Rapid Portable Electrical Biosensing Design With Dielectrophoresis and Its Application for Cardiac Biomarker Detection
顯示簡易詮釋資料 統計資料| 欄位 | 內容 | 語言 |
|---|---|---|
dc.citation.epage |
8989 | en_US |
dc.citation.issue |
16 | en_US |
dc.citation.spage |
8981 | en_US |
dc.citation.volume |
20 | en_US |
dc.citation.woscount |
0 | en_US |
dc.contributor.author |
Chen, Chun-Chi | en_US |
dc.contributor.author |
Chan, Yun-Sheng | en_US |
dc.contributor.author |
Fang, Yu-Hao | en_US |
dc.contributor.author |
Wang, Yun-Ming | en_US |
dc.contributor.author |
Lee, Chen-Yi | en_US |
dc.contributor.department |
生物科技學系 | zh_TW |
dc.contributor.department |
電子工程學系及電子研究所 | zh_TW |
dc.contributor.department |
Department of Biological Science and Technology | en_US |
dc.contributor.department |
Department of Electronics Engineering and Institute of Electronics | en_US |
dc.date.accessioned |
2020-10-05T02:01:06Z | |
dc.date.available |
2020-10-05T02:01:06Z | |
dc.date.issued |
2020-08-01 | en_US |
dc.description.abstract |
With the growing number of new biomarkers discovered for clinic diagnosis, there is an emerging need for the development of biosensing device. One important application of the biosensor in disease diagnosis is to detect biomarkers and the biosensor should be sensitive enough to measure the small variation in biomarker concentration. Besides, to support the point-of-care diagnostics that can facilitate emergency diagnosis and disease screening, it is important that the sensing device needs to be portable, user-friendly, and support the rapid biomarker detection. In this work, we propose a rapid portable sensing device by utilizing dielectrophoresis force to accelerate the centralization of the biomarkers and the comparator-based operational amplifier to enlarge small signal response. Furthermore, the testing results demonstrate the proposed biosensor can detect subtle admittance variation corresponding to the concentration change of the biomarkers in short assay time. The sensing device is able to detect cardiac troponins level with the limit of detection at 0.1 ng mL(-1), that can satisfy the actual requirements for diagnosis of cardiovascular disease. The proposed biosensor can provide an easy-to-use, cost-effective, and rapid biomarker detection that can be applied to clinical applications to support point-of-care diagnostics. | en_US |
dc.identifier.doi |
10.1109/JSEN.2020.2988884 | en_US |
dc.identifier.issn |
1530-437X | en_US |
dc.identifier.journal |
IEEE SENSORS JOURNAL | en_US |
dc.identifier.uri |
http://dx.doi.org/10.1109/JSEN.2020.2988884 | en_US |
dc.identifier.uri |
https://ir.lib.nycu.edu.tw/handle/11536/155145 | |
dc.identifier.wosnumber |
WOS:000550685000010 | en_US |
dc.language.iso |
en_US | en_US |
dc.subject |
Biomarker | en_US |
dc.subject |
biosensor | en_US |
dc.subject |
comparator-based op-amp | en_US |
dc.subject |
dielectrophoresis | en_US |
dc.subject |
point-of-care | en_US |
dc.title |
Rapid Portable Electrical Biosensing Design With Dielectrophoresis and Its Application for Cardiac Biomarker Detection | en_US |
dc.type |
Article | en_US |
檔案
本條目沒有附加檔案。
所屬典藏: 期刊論文
引用此條目
Chen, C.-C., Chan, Y.-S., Fang, Y.-H., Wang, Y.-M., & Lee, C.-Y. (2020). Rapid Portable Electrical Biosensing Design With Dielectrophoresis and Its Application for Cardiac Biomarker Detection. IEEE SENSORS JOURNAL, 20(16), 8981–8989. https://doi.org/10.1109/JSEN.2020.2988884
引用文字依 APA 第 7 版格式自動產生,使用前請再核對。
分享