<link rel="stylesheet" href="styles.cf390c6bfbe7555f.css"> Skip to main content

PHOTOPLETHYSMOGRAPHY SIGNALS PROCESSING USING POLYNOMIAL PROFILE FITTING FOR MEASURING THE BLOOD FLOW VOLUME IN ARTERIOVENOUS FISTULA

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

DOI

Abstract

In this work, a new photoplethysmography (PPG) signal processing methodology using polynomial profile fitting for measuring blood flow volume (BFV) in arteriovenous fistula (AVF) is proposed. After calibration, the experiment results using proposed method shows higher correlation (R=0.7883) and lower error (RMSE=109 ml/min) compared to the ones using conventional methods (R=0.3212, RMSE=168 ml/min).

Description

Keywords

Citation

Endorsement

Review

Supplemented By

Referenced By