3D pH MEASUREMENT FOR CELL CULTURE USING CT SCAN
Loading...
Date
Journal Title
Journal ISSN
Volume Title
Publisher
DOI
Abstract
In cell culture, the pH of culture liquid is one of the most important conditions. However, culture liquid may have pH-gradient due to the activity of cells and change of the environment. In order to confirm the existence of pH-gradient, we propose a method for 3D pH measurement using the color change of phenol red. Using the principle of CT scan, we can measure pH-gradient without contacting culture liquid, and thus, decrease the possibility of contamination. In this paper, we consider the influence of light refraction by a cylindrical vessel, and show that we can measure local pH of culture liquid by light absorbance.