A Fully-Integrated Charge Pump for Self-Powered Implantable Retinal Prostheses
Loading...
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
DOI
Abstract
This paper presents a compact fully-integrated charge pump based voltage doubler for self-powered implantable retinal prostheses. The target input voltage is 0.55 V to meet the output voltage of on-chip photovoltaic cells, and the target output is around 1 V to provide sufficient current for neuron stimulation. Several techniques are adopted in the proposed design, including gate-boosting circuits to enhance current driving capability, a time-interleaving technique to reduce the size of the output capacitor, and a non-overlapping clock generator to suppress the shoot-through current. The proposed charge pump was implemented in 0.18 mu m CMOS process with core chip area of 700 mu m x 280 mu m. The measurement results demonstrate that the proposed charge pump is capable of providing around 1 V output from an input voltage of 0.55 V with 500 mu A loading current. The peak power efficiency is 73.4 % and the total area is 700 mu m x 280 mu m.