The Influence of Relative Humidity on Nanoparticle Concentration and Particle Mass Distribution Measurements by the MOUDI

檔案

日期

2011

類型

期刊名稱

AEROSOL SCIENCE AND TECHNOLOGY

卷期與頁次

第 45 卷第 5 期,頁 596–603

ISSN

0278-6826

語言

英文

摘要

A humidity control system was operated upstream of two collocated-MOUDIs (micro-orifice uniform deposit impactors) for sampling ambient aerosol particles. One MOUDI used silicone-grease-coated aluminum foils (ALs) as the impaction substrates and was considered as the reference impactor, while the other used uncoated ALs or uncoated Teflon filters (TFs) as the impaction substrates for quantifying the effect of different relative humidities (RHs) and impaction substrates on the PM(0.1) concentrations and mass distributions of ambient PMs. Test results showed that decreasing RH in general increased particle bounce from uncoated substrates with the bounce from uncoated ALs being more severe than that from uncoated TFs. Particle bounce did not influence the overall mass distribution of ambient fine particles when RH ranged between 40% and 80%, whereas it led to undersampling of particles greater than 2.5 mu m in aerodynamic diameter severely. Oversampling of PM(0.1) occurred by as much as 95%-180% or 25%-55% when the MOUDI used uncoated ALs or TFs, respectively, as RH was reduced from 50% to 25%. Particle bounce was found to be negligible, and PM(0.1) and PM(2.5) could be sampled accurately with less than 5% error at the RH of 75%-80% or 65%-80% when uncoated ALs or TFs were used, respectively.

所屬典藏

期刊論文
顯示完整詮釋資料 統計資料

引用此條目

Chen, S.-C., Tsai, C.-J., Chen, H.-D., Huang, C.-Y., & Roam, G.-D. (2011). The Influence of Relative Humidity on Nanoparticle Concentration and Particle Mass Distribution Measurements by the MOUDI. AEROSOL SCIENCE AND TECHNOLOGY, 45(5), 596–603. https://doi.org/10.1080/02786826.2010.551557

BibTeX RIS

引用文字依 APA 第 7 版格式自動產生,使用前請再核對。

分享

永久網址:https://ir.lib.nycu.edu.tw/handle/11536/26074