An electrochemical dopamine sensor with CMOS detection circuit

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

DOI

10.1109/ICSENS.2007.4388686

Abstract

This paper presents the use of interdigitated microelectrodes as the platform for electrochemical sensing of the neurotransmitter dopamine. Gold electrodes with gaps ranging between 3 gm to 7 gm are fabricated by the lift-off technique. The reduction and oxidation potentials for dopamine are determined by voltammetry using the fabricated microelectrodes. A CMOS sensing circuit is used to amplify the electrochemical current produced by the working electrodes. Experimental result shows that for a dopamine concentration of 10 mu M the measured current value is 11 pA.