Title: ATMOSPHERIC AEROSOL SAMPLING BY AN ANNULAR DENUDER SYSTEM AND A HIGH-VOLUME PM(10) SAMPLER
Authors: TSAI, CJ
HUANG, HY
交大名義發表
環境工程研究所
National Chiao Tung University
Institute of Environmental Engineering
Issue Date: 1995
Abstract: An Annular Denuder System (ADS) was used to study sampling errors of total PM(10) and soluble ionic species concentrations determined by the high volume PM(10) (HVPM(10)) sampler. Experimental results show that in the HVPM(10) sampler, more alkaline glass fiber filters collect more sulfate and nitrate but less ammonium ion than quartz filters. However, total PM(10) concentrations do not seem to depend on the types of filters. Artifact sulfate concentrations in glass fiber filters of the HVPM(10) sampler are found to exceed those predicted by the Coutant's model. It is suggested that artifact sulfate and nitrate not only occur within filters but also in particle deposits. Chloride ion concentrations in the filters of the HVPM(10) sampler are underestimated significantly due to evaporative loss of ammonium chloride.
URI: http://hdl.handle.net/11536/2162
ISSN: 0160-4120
Journal: ENVIRONMENT INTERNATIONAL
Volume: 21
Issue: 3
Begin Page: 283
End Page: 291