Design and characterization of high-temperature-resolution thermometer
檔案
- 000235692100056.pdf (735.4 KB, Adobe PDF)
日期
2006-02-01
類型
系所
期刊名稱
JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS BRIEF COMMUNICATIONS & REVIEW PAPERS
卷期與頁次
第 45 卷第 2A 期,頁 896–899
ISSN
0021-4922
語言
英文
摘要
Thermochromic liquid crystals (TLCs) can serve as a light valve when it is carefully controlled by temperature. With a photodiode used in light transmissive intensity electric output voltage extraction, there is a progressive increase in output voltage from 19.331 to 35.800 degrees C, and the resolution is 0.0001482 degrees C, which decreases progressively to 0.03086 degrees C. If we design a linear integrated circuit with many very small sections for converting nonlinear extraction voltage to its relative temperature automatically, this system can serve as a thermometer. The whole system can be made portable for radiation temperature measurement.
所屬典藏
期刊論文引用此條目
Shie, J., & Hwang, T. (2006). Design and characterization of high-temperature-resolution thermometer. JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS BRIEF COMMUNICATIONS & REVIEW PAPERS, 45(2A), 896–899. https://doi.org/10.1143/JJAP.45.896
引用文字依 APA 第 7 版格式自動產生,使用前請再核對。
分享
永久網址:https://ir.lib.nycu.edu.tw/handle/11536/12697