Use of porous foam as the substrate of an impactor for respirable aerosol sampling

檔案

日期

2005-11-01

類型

期刊名稱

JOURNAL OF AEROSOL SCIENCE

卷期與頁次

第 36 卷第 11 期,頁 1373–1386

ISSN

0021-8502

語言

英文

摘要

The effect of a porous foam substrate on the penetration characteristics of an impactor with a single round nozzle was investigated experimentally, with liquid and solid particles in the size range of 1-10 mu m. The experimental results show that the cut-off aerodynamic diameter of the impactor with porous foam substrate is lower than that with porous metal substrate due to more air penetration into the porous foam; in addition, the impactor penetration curve is less sharp. A numerical simulation of the penetration curve for the impactor with porous foam substrate was in good agreement with the experimental data. The numerical method was then used to find the best sampling flow rate for respirable aerosol sampling. Results show that when the nozzle diameter is 0.36 cm and the sampling rate is 1.91/min, the penetration curve of the impactor with the porous foam is very close to the criteria for respirable aerosol sampling. (C) 2005 Elsevier Ltd. All rights reserved.

所屬典藏

期刊論文
顯示完整詮釋資料 統計資料

引用此條目

Huang, C., Chang, C., Chang, S., Tsai, C., Shih, T., & Tang, D. (2005). Use of porous foam as the substrate of an impactor for respirable aerosol sampling. JOURNAL OF AEROSOL SCIENCE, 36(11), 1373–1386. https://doi.org/10.1016/j.jaerosci.2005.03.008

BibTeX RIS

引用文字依 APA 第 7 版格式自動產生,使用前請再核對。

分享

永久網址:https://ir.lib.nycu.edu.tw/handle/11536/13102