An evaluation of clinical accuracy of the EasyTouch blood uric acid self-monitoring system

檔案

日期

2005-03-01

類型

期刊名稱

CLINICAL BIOCHEMISTRY

卷期與頁次

第 38 卷第 3 期,頁 278–281

ISSN

0009-9120

語言

英文

摘要

Objectives: Self-monitoring blood uric acid device is an important tool for patients to efficiently monitor their blood uric acid concentrations. The objective of the present study was to evaluate the accuracy of EasyTouch uric acid monitoring system. Design and methods: Capillary blood uric acid concentrations measured using EasyTouch and the reference values obtained from COBAS MIRA were performed in the Department of Laboratory Medicine, Wei-Gong Memorial Hospital. Results were evaluated using (1) linear regression analysis, (2) percentage of readings within a defined range of deviation from the reference value, and (3) coefficients of variation (CVs) calculated from 60 measurements in series. Results: The window of the 177 EasyTouch readings covered a wide range from 0.1785 to 0.6367 mmol/L (3-10.7 mg/dL). Linear regression analysis yielded a regression slope of 0.975, an intercept of 0.0118 mmol/L and an R-2 of 0.8966. Of the EasyTouch readings, 64 (36.2%), 61 (34.5%), 34 (19.2%), 9 (5.08%), and 9 (5.08%) were within the intervals of <5%, 5-10%, 10-15%, 15-17%, and >17%, respectively, of the reference values. Further analysis for the performance of each lot of strips showed that both coefficients of correlation and the percentages of readings within the CLIA's criterion (+/-17%), respectively, were in a narrow range from 0.8777 to 0.9541 and from 92.1% to 100%. The CVs for the seven lots of strips (lot I to lot 7) ranged from 2.93% to 6.33%, 3.2% to 5.9%, 3.64% to 7.0%, 2.84% to 7.6%, 2.68% to 5.42%, 3.03% to 6.93%, and 3.18% to 5.17%, respectively. Conclusion: EasyTouch is an acceptable diagnostic device which provides accurate uric acid measurements. (C) 2004 The Canadian Society of Clinical Chemists. All rights reserved.

所屬典藏

期刊論文
顯示完整詮釋資料 統計資料

引用此條目

Dai, K., Tai, D., Ho, P., Chen, C., Peng, W., Chen, S., Hsu, C., Liu, Y., Hsieh, H., & Mao, S. (2005). An evaluation of clinical accuracy of the EasyTouch blood uric acid self-monitoring system. CLINICAL BIOCHEMISTRY, 38(3), 278–281. https://doi.org/10.1016/j.clinbiochem.2004.11.005

BibTeX RIS

引用文字依 APA 第 7 版格式自動產生,使用前請再核對。

分享

永久網址:https://ir.lib.nycu.edu.tw/handle/11536/13948