A MEMS µ-Preconcentrator Employing a Carbon Molecular Sieve Membrane for Highly Volatile Organic Compound Sampling

檔案

本條目沒有附加檔案。

至出版者網站查看(DOI) ›

日期

2021-05-07

類型

期刊名稱

Chemosensors

卷期與頁次

第 9 卷第 5 期,頁 104

ISSN

2227-9040

語言

英文

摘要

This paper presents the synthesis and evaluation of a carbon molecular sieve membrane (CMSM) grown inside a MEMS-fabricated μ-preconcentrator for sampling highly volatile organic compounds. An array of µ-pillars measuring 100 µm in diameter and 250 µm in height were fabricated inside a microfluidic channel to increase the attaching surface for the CMSM. The surface area of the CMSM was measured as high as 899 m2/g. A GC peak amplification factor >2 × 104 was demonstrated with gaseous ethyl acetate. Up to 1.4 L of gaseous ethanol at the 100 ppb level could be concentrated without exceeding the capacity of this microchip device. Sharp desorption chromatographic peaks (

所屬典藏

期刊論文
顯示完整詮釋資料 統計資料

引用此條目

Kuo, H.-Y., Cheng, W.-R., Wu, T.-H., Sheen, H.-J., Wang, C.-C., & Lu, C.-J. (2021). A MEMS µ-Preconcentrator Employing a Carbon Molecular Sieve Membrane for Highly Volatile Organic Compound Sampling. Chemosensors, 9(5), 104. https://doi.org/10.3390/chemosensors9050104