An Image Quality Comparison Study Between Homemade and Commercial Dental Cone-Beam CT Systems

檔案

本條目沒有附加檔案。

至出版者網站查看(DOI) ›

日期

2021-11-15

類型

期刊名稱

Journal of Medical and Biological Engineering

卷期與頁次

第 41 卷第 6 期,頁 870–880

ISSN

1609-0985
2199-4757

語言

英文

摘要

Purpose Cone-beam computed tomography (CBCT) has been widely applied in dental and maxillofacial imaging. Several dental CBCT systems have been recently developed in order to improve the performance. This study aimed to evaluate the image quality of our prototype (YMU-DENT-P001) and compare with a commercial POYE Expert 3DS dental CBCT system (system A). Methods The Micro-CT Contrast Scale, Micro-CT Water and Micro-CT HA phantoms were used to evaluate the two CBCT systems in terms of contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), uniformity (U), distortion, and linearity in the relationship between image intensity and calcium hydroxyapatite concentration. We also fabricated a proprietary thin-wire phantom to evaluate full width at half maximum (FWHM) spatial resolution. Both CBCT systems used the same exposure protocol, and data analysis was performed in accordance with ISO standards using a proprietary image analysis platform. Results The SNR of our prototype system was nearly five times higher than that of system A (prototype: 159.85 ± 3.88; A: 35.42 ± 0.61; p < 0.05) and the CNR was three times higher (prototype: 329.39 ± 5.55; A: 100.29 ± 2.31; p < 0.05). The spatial resolution of the prototype (0.2446 mm) greatly exceeded that of system A (0.5179 mm) and image distortion was lower (prototype: 0.03 mm; system A: 0.285 mm). Little difference was observed between the two systems in terms of the linear relationship between bone mineral density (BMD) and image intensity. Conclusions Within the scope of this study, our prototype YMU-DENT-P001 outperformed system A in terms of spatial resolution, SNR, CNR, and image distortion.

所屬典藏

期刊論文
顯示完整詮釋資料 統計資料

引用此條目

Tran, T. T. N., Jin, D. S.-C., Shih, K.-L., Hsu, M.-L., & Chen, J.-C. (2021). An Image Quality Comparison Study Between Homemade and Commercial Dental Cone-Beam CT Systems. Journal of Medical and Biological Engineering, 41(6), 870–880. https://doi.org/10.1007/s40846-021-00663-7

BibTeX RIS

引用文字依 APA 第 7 版格式自動產生,使用前請再核對。

分享

永久網址:https://ir.lib.nycu.edu.tw/handle/11536/170359