ECIS Based Electric Fence Method for Measurement of Human Keratinocyte Migration on Different Substrates
檔案
本條目沒有附加檔案。
日期
2022-05-03
類型
系所
期刊名稱
Biosensors
卷期與頁次
第 12 卷第 5 期,頁 293
ISSN
2079-6374
語言
英文
摘要
Electric Cell-substrate Impedance Sensing (ECIS) is an impedance-based, real-time, and label-free measuring system for monitoring cellular activities in tissue culture. Previously, ECIS wound healing assay has been used to wound cells with high electric current and monitor the subsequent cell migration. In this study, we applied ECIS electric fence (EF) method, an alternative to electrical wounding, to assess the effects of different surface coatings on human keratinocyte (HaCaT) migration. The EF prevents inoculated cells from attaching or migrating to the fenced electrode surface while maintaining the integrity of the surface coating. After the EF is turned off, cells migrate into the cell-free area, and the increase in measured impedance is monitored. We cultured HaCaT cells on gold electrodes without coating or coated with poly-L-lysin (PLL), poly-D-lysine (PDL), or type-I collagen. We quantified migration rates according to the different slopes in the impedance time series. It was observed that either poly-L-lysine (PLL) or poly-D-lysine (PDL) limits cell adhesion and migration rates. Furthermore, the surface charge of the coated substrate in the culture condition positively correlates with the cell adhesion and migration process. Our results indicate that the EF method is useful for determining cell migration rates on specific surface coatings.
所屬典藏
期刊論文引用此條目
Hung, Y.-H., Chiu, W.-C., Fuh, S.-R., Lai, Y.-T., Tung, T.-H., Huang, C.-C., & Lo, C.-M. (2022). ECIS Based Electric Fence Method for Measurement of Human Keratinocyte Migration on Different Substrates. Biosensors, 12(5), 293. https://doi.org/10.3390/bios12050293
引用文字依 APA 第 7 版格式自動產生,使用前請再核對。
分享
永久網址:https://ir.lib.nycu.edu.tw/handle/11536/173299