Direct Mass Spectrometric Analysis of Semivolatiles Derived from Real Samples at Atmospheric Pressure

檔案

本條目沒有附加檔案。

至出版者網站查看(DOI) ›

日期

2022-03-18

類型

期刊名稱

ACS Omega

卷期與頁次

第 7 卷第 12 期,頁 10255–10261

ISSN

2470-1343

語言

英文

摘要

This study demonstrated a facile ionization method with the use of real samples for the ionization of their main compositions at ambient conditions for mass spectrometric analysis. Analyte ions derived from the real samples were readily observed in the mass spectrum when placing the samples close (≤1 mm) to the inlet of the mass spectrometer applied with a high voltage. No additional accessories such as an ionization emitter, a plasma generator, or a high voltage power supply were required for this approach. Ionization of semivolatiles derived from the samples occurred between the samples and the inlet of the mass spectrometer presumably owing to the dielectric breakdown induced by the electric field provided by the mass spectrometer. Real samples including plants, medicine tablets, and gloves with contaminants were used as the model samples. The putative ionization mechanisms are also discussed in this study.

所屬典藏

期刊論文
顯示完整詮釋資料 統計資料

引用此條目

Huang, D.-Y., Tsai, J.-J., & Chen, Y.-C. (2022). Direct Mass Spectrometric Analysis of Semivolatiles Derived from Real Samples at Atmospheric Pressure. ACS Omega, 7(12), 10255–10261. https://doi.org/10.1021/acsomega.1c06869