Analytical validation of an inductively coupled plasma mass spectrometry method for urinary iodine concentration measurements in Taiwan

檔案

本條目沒有附加檔案。

至出版者網站查看(DOI) ›

日期

2023-03-04

類型

期刊名稱

Journal of the Formosan Medical Association

卷期與頁次

第 122 卷第 8 期,頁 757–765

ISSN

0929-6646
1876-0821

語言

英文

摘要

BACKGROUND: Urinary iodine concentration (UIC) measured by Sandell-Kolthoff spectrophotometric method has been used in the Nutrition and Health Surveys in Taiwan but this method is time consuming and produces toxic waste from arsenic trioxide. The aim of this study was to develop and validate an inductively coupled plasma mass spectrometry (ICP-MS) system to determine UIC in Taiwan. METHODS: Samples and iodine calibrators were diluted 100-fold into an aqueous solution containing Triton X-100, 0.5% ammonia solution, and tellurium (128Te) as an internal standard. Digestion prior to analysis was not necessary. Precision, accuracy, serial dilution, and recovery tests were performed. A total of 1243 urine samples covering a wide range of iodine concentrations were measured by both Sandell-Kolthoff method and ICP-MS. Passing-Bablok regression and Bland-Altman plots were used to compare values across methods. RESULTS: The limit for detection and quantification by ICP-MS was 0.95 μg/L and 2.85 μg/L, respectively. The intra-assay and inter-assay coefficients were

所屬典藏

期刊論文
顯示完整詮釋資料 統計資料

引用此條目

Huang, C.-J., Lee, L.-H., Cheng, C.-P., Chen, H.-S., Hwu, C.-M., Tang, K.-T., Shih, C.-W., Yeh, C.-C., Wang, F.-F., & Yang, C.-C. (2023). Analytical validation of an inductively coupled plasma mass spectrometry method for urinary iodine concentration measurements in Taiwan. Journal of the Formosan Medical Association, 122(8), 757–765. https://doi.org/10.1016/j.jfma.2023.02.010

BibTeX RIS

引用文字依 APA 第 7 版格式自動產生,使用前請再核對。

分享

永久網址:https://ir.lib.nycu.edu.tw/handle/11536/178485