Biomechanical Effects of a Novel Pedicle Screw W-Type Rod Fixation for Lumbar Spondylolysis: A Finite Element Analysis

檔案

本條目沒有附加檔案。

至出版者網站查看(DOI) ›

日期

2023-04-07

類型

系所

期刊名稱

Bioengineering

卷期與頁次

第 10 卷第 4 期,頁 451

ISSN

2306-5354

語言

英文

摘要

Lumbar spondylolysis involves anatomical defects of the pars interarticularis, which causes instability during motion. The instability can be addressed through instrumentation with posterolateral fusion (PLF). We developed a novel pedicle screw W-type rod fixation system and evaluated its biomechanical effects in comparison with PLF and Dynesys stabilization for lumbar spondylolysis via finite element (FE) analysis. A validated lumbar spine model was built using ANSYS 14.5 software. Five FE models were established simulating the intact L1-L5 lumbar spine (INT), bilateral pars defect (Bipars), bilateral pars defect with PLF (Bipars_PLF), Dynesys stabilization (Bipars_Dyn), and W-type rod fixation (Bipars_Wtyp). The range of motion (ROM) of the affected segment, the disc stress (DS), and the facet contact force (FCF) of the cranial segment were compared. In the Bipars model, ROM increased in extension and rotation. Compared with the INT model, Bipars_PLF and Bipars_Dyn exhibited remarkably lower ROMs for the affected segment and imposed greater DS and FCF in the cranial segment. Bipars_Wtyp preserved more ROM and generated lower stress at the cranial segment than Bipars_PLF or Bipars_Dyn. The injury model indicates that this novel pedicle screw W-type rod for spondylolysis fixation could return ROM, DS, and FCF to levels similar to preinjury.

所屬典藏

期刊論文
顯示完整詮釋資料 統計資料

引用此條目

Pan, J.-H., Chen, C.-S., Liu, C.-L., & Chou, P.-H. (2023). Biomechanical Effects of a Novel Pedicle Screw W-Type Rod Fixation for Lumbar Spondylolysis: A Finite Element Analysis. Bioengineering, 10(4), 451. https://doi.org/10.3390/bioengineering10040451

BibTeX RIS

引用文字依 APA 第 7 版格式自動產生,使用前請再核對。

分享

永久網址:https://ir.lib.nycu.edu.tw/handle/11536/178581