Evaluating Interface Pressure Distribution and Range of Motion in a Mesh-Structure 3D-Printed Lumbar Brace
檔案
本條目沒有附加檔案。
日期
2026-08-12
類型
系所
期刊名稱
Bioengineering
卷期與頁次
第 13 卷第 8 期,頁 910
ISSN
2306-5354
語言
英文
摘要
Conventional spinal braces often suffer from poor fit and low breathability, with most 3D-printing research focusing on scoliosis rather than general support. This study evaluated the biomechanical performance and range of motion (ROM) restriction of a semi-custom 3D-printed brace (3DP) compared with two traditional braces (Medi-H, Medi-S). Sixteen participants performed static standing, flexion, extension, and pick-up tasks while wearing each brace. Interface pressure was recorded across seven anatomical regions, and thoracic (C7-T8) and lumbar (T8-L3) ROM were assessed simultaneously using repeated-measures ANOVA. Results showed that 3DP and Medi-H provided significantly greater overall pressures than Medi-S. During flexion and pick-up tasks, 3DP delivered enhanced support to the middle lumbar region and erector spinae, whereas Medi-H concentrated greater pressure on the spine, particularly during extension. Thoracic and lumbar ROM were restricted comparably across all braces (20-30°). In conclusion, 3DP achieves an equivalent restriction of sagittal motion to traditional braces while offering superior regional support to the upper and middle lumbar regions.
所屬典藏
期刊論文引用此條目
Hsu, H.-C., Chen, P.-Y., & Chen, C.-S. (2026). Evaluating Interface Pressure Distribution and Range of Motion in a Mesh-Structure 3D-Printed Lumbar Brace. Bioengineering, 13(8), 910. https://doi.org/10.3390/bioengineering13080910
引用文字依 APA 第 7 版格式自動產生,使用前請再核對。
分享
永久網址:https://ir.lib.nycu.edu.tw/handle/11536/196111