Agreement and Systematic Differences Between Intraoperative ROSA Hip-Knee-Ankle Measurements and Standing Radiographic Alignment After Robotic-Assisted Total Knee Arthroplasty: A Retrospective Agreement Study
檔案
-
10.3390_jcm15197432.pdf
(1.69 MB, Adobe PDF)
出版社正式版,依 CC BY 授權轉載,來源 https://doi.org/10.3390/jcm15197432
日期
2026-09-24
作者
類型
系所
期刊名稱
Journal of Clinical Medicine
卷期與頁次
第 15 卷第 19 期,頁 7432
ISSN
2077-0383
語言
英文
摘要
Background/Objectives: Intraoperative robotic hip-knee-ankle angle (HKA) measurements are obtained with the patient supine under operative conditions and may not directly represent postoperative standing weight-bearing alignment. This study quantified the correspondence and systematic differences between intraoperative ROSA HKA and standing radiographic HKA after robotic-assisted total knee arthroplasty. Methods: This retrospective single-center study included 208 knees from 178 consecutive patients, including 30 bilateral patients. The primary comparison was post-trial-component intraoperative ROSA HKA versus postoperative standing radiographic HKA; the secondary comparison was pre-cut intraoperative ROSA HKA versus preoperative standing radiographic HKA. Patient-cluster bootstrap confidence intervals used 10,000 resamples. Agreement was evaluated using Bland-Altman analysis, absolute-agreement ICC(A,1), and weighted kappa; Spearman ρ described association. Results: For the primary comparison, mean bias (ROSA minus radiographic HKA) was -1.75° (95% CI -2.08 to -1.43°), conventional limits of agreement were -6.36° to 2.85°, and ICC(A,1) was 0.472 (95% CI 0.379 to 0.557). Differences were within 3° in 70.2% of knees and exceeded 3° in 29.8%; proportional bias was present (slope -0.569, p < 0.001). The secondary comparison showed bias 0.97° and ICC(A,1) 0.956. Exact femoral and tibial size agreement was 42.3% and 80.8%, respectively. The standardized surgeon-entered 10 mm starting value was retained in 57.7% of knees. These systematic and angle-dependent differences may partly reflect the differing loading, positioning, and measurement conditions between the intraoperative trial construct and postoperative standing radiography. Conclusions: Post-trial-component ROSA HKA showed systematic and angle-dependent differences from postoperative standing HKA and should not be considered interchangeable with, or a fixed patient-level predictor of, postoperative standing alignment.
所屬典藏
期刊論文引用此條目
Tseng, W.-C., Lin, C.-H., Lin, Y.-T., Wang, C.-C., Tu, K.-C., Chen, K.-H., Lee, C.-H., Ho, C.-J., & Tzeng, C.-Y. (2026). Agreement and Systematic Differences Between Intraoperative ROSA Hip-Knee-Ankle Measurements and Standing Radiographic Alignment After Robotic-Assisted Total Knee Arthroplasty: A Retrospective Agreement Study. Journal of Clinical Medicine, 15(19), 7432. https://doi.org/10.3390/jcm15197432
引用文字依 APA 第 7 版格式自動產生,使用前請再核對。
分享
永久網址:https://ir.lib.nycu.edu.tw/handle/11536/196674