Title: | A +/- 6ms-Accuracy, 0.68mm(2) and 2.21 mu W QRS Detection ASIC |
Authors: | Wang, Hui-Min Lai, You-Liang Hou, Mark C. Lin, Shih-Hsiang Yen, Brad S. Huang, Yu-Chieh Chou, Lei-Chun Hsu, Shao-You Huang, Sheng-Chieh Jan, Ming-Yie 電機工程學系 Department of Electrical and Computer Engineering |
Issue Date: | 2010 |
Abstract: | Healthcare issues arose from population aging. Meanwhile, electrocardiogram (ECG) is a powerful measurement tool. The first step of ECG is to detect QRS complexes. A state-of-the-art QRS detection algorithm was modified and implemented. By the dedicated architecture design, the novel ASIC is proposed with 2.21 mu W power consumption and 0.68mm(2) core area. It is the smallest QRS detection ASIC so far in the world. In addition, the positive prediction of the ASIC is 99.36% based on the MIT/BIH arrhythmia database certification. |
URI: | http://hdl.handle.net/11536/26354 |
ISBN: | 978-1-4244-5309-2 |
ISSN: | 0271-4302 |
Journal: | 2010 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS |
Begin Page: | 1372 |
End Page: | 1375 |
Appears in Collections: | Conferences Paper |