標題: | Study of in vitro Diagnostic Preconcentration Technique via Surface Acoustic Wave Modified With the 54 MHz Communication Chip |
作者: | Chien, Ming-Hung Chiang, S. H. Chuang, Ching-Cheng Chen, Jung-Chih 分子醫學與生物工程研究所 Institute of Molecular Medicine and Bioengineering |
關鍵字: | Specimen preconcentration;Biosensor;Surface acoustic wave;Microchannel;Acoustofluidics |
公開日期: | 1-一月-2018 |
摘要: | Particle separation and specimen preconcentration techniques are ones of the essential but time-consuming steps in the beginning of many biomedical sensing experiments or biosensor design. Hence, the study demonstrates a simple microfluidic system combined with surface acoustic wave (SAW) device, hoping to help researchers could spend less time on this routine process. A detachable and reusable acrylic microchannel is placed on the customized SAW chip which releases a 54 MHz SAW wave to separate 10 mu m particles from the others (1 mu m & 6 mu m) under the cut-off input power 23.4 dBm verified in the study and about 90% separation efficiency in average. |
URI: | http://dx.doi.org/10.1109/IS3C.2018.00105 http://hdl.handle.net/11536/151712 |
ISBN: | 978-1-5386-7036-1 |
ISSN: | 2476-1052 |
DOI: | 10.1109/IS3C.2018.00105 |
期刊: | 2018 INTERNATIONAL SYMPOSIUM ON COMPUTER, CONSUMER AND CONTROL (IS3C 2018) |
起始頁: | 392 |
結束頁: | 395 |
顯示於類別: | 會議論文 |