A High-Sensitivity CMOS-Compatible Biosensing System Based on Absorption Photometry

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

DOI

10.1109/JSEN.2008.2011079

Abstract

An optical biosensing system based on a standard CMOS technology and absorption photometry is proposed. Within the compact prototype, a CMOS sensor consisting a P(+)/N(well) finger photodiode and a transimpedance amplifier was implemented in a standard 0.35-mu m CMOS technology. The ABTS/H(2)O(2)/HRP method was adopted to be coupled with various analytes of interest, such as glucose and histamine, which can serve as the basis for medical diagnosis. The experimental results demonstrate that the minimum concentrations successfully detected for H(2)O(2), glucose, and histamine are 1 mu M, 1 mu M, and 10 mu M, respectively. The detection limits are at least one order of magnitude better than those of reported CMOS biosensors, and are even comparable to those of a commercial spectrophotometer.