標題: Non-uniform Dose Measurement Based on Two-Stage Bleaching Integrated with Thermo-luminescence and Optically Stimulated Luminescence Dosimeter
作者: Hsieh, Chi-Wen
Lin, Yi-Hong
物理研究所
Institute of Physics
關鍵字: thermoluminescence;optically stimulated luminescence;optical bleaching
公開日期: 2013
摘要: Thermo luminescence dosimeter and optically stimulated luminescence dosimeter are the most popular dosimeters for personal and environmental radiation monitoring. Basically, these dosimeters are both accumulated dosimeters so that they cannot precisely measure dynamic doses, i.e. there is no any information of time resolving. We developed an algorithm to solve the problem that can depress the uncertain measurement by using partially bleaching techniques; the two-stage bleaching method under the fixed bleaching intensity condition. To analyze and compare the difference of ionizing radiation dose between two dosimeters can partially recall the non-uniformly distributed radiation doses. The Gaussian distribution of dose is considered for the simulation and the results indicated that error of the dose rate could be depressed to 15% compared to the traditional method.