標題: USING THE TIME-DOMAIN CHARACTERIZATION FOR ESTIMATION CONTINUOUS BLOOD PRESSURE VIA NEURAL NETWORK METHOD
作者: Chao, Paul C. -P.
Tu, Tse-Yi
電控工程研究所
Institute of Electrical and Control Engineering
公開日期: 1-一月-2017
摘要: The new method with back-propagation neural network is expected to be capable of continuous measurement of blood pressures with noninvasive, cuffless strain blood pressure sensor. The eight time-domain characterizations estimate systolic blood pressure and diastolic blood pressure via BPNN leading to a satisfactory accuracy of the BP sensor. The BP sensor is used on human wrist to collect the continuously pulse signal for measuring blood pressures. To assist the sensor, a readout circuit is devised with a Wheatstone bridge, amplifier, filter, and a digital signal processor. The results of SBP and DBP are 4.27 +/- 4.98 mmHg and 3.86 +/- 5.35 mmHg, respectively. The errors of blood pressure pass the criteria for Association for the Advancement of Medical Instrumentation (AAMI) method 2 and the British Hypertension Society (BHS) Grade B.
URI: http://hdl.handle.net/11536/146874
期刊: PROCEEDINGS OF THE ASME 26TH ANNUAL CONFERENCE ON INFORMATION STORAGE AND PROCESSING SYSTEMS, 2017
顯示於類別:會議論文