標題: A Fully Integrated Closed-Loop Neuromodulation SoC with Wireless Power and Bidirectional Data Telemetry for Real-Time Human Epileptic Seizure Control
作者: Cheng, Cheng-Hsiang
Tsai, Ping-Yuan
Yang, Tzu-Yi
Cheng, Wan-Hsueh
Yen, Ting-Yang
Luo, Zhicong
Qian, Xin-Hong
Chen, Zhi-Xin
Lin, Tsu-Han
Chen, Wei-Hong
Chen, Wei-Ming
Liang, Sheng-Fu
Shaw, Fu-Zen
Chang, Cheng-Siu
Shih, Fu-Yuan
Hsin, Yue-Loong
Lee, Chen-Yi
Ker, Ming-Dou
Wu, Chung-Yu
交大名義發表
National Chiao Tung University
公開日期: 1-一月-2017
摘要: This paper presents a 16-channel closed-loop neuromodulation SoC for human seizure control. The SoC includes a 16-ch signal acquisition unit, a bio-signal processor, a 16-ch adaptive stimulator, and wireless telemetry. The signal acquisition unit achieves 3.78 NEF and shares electrodes with stimulator. The seizure detection latency is 0.76s and delivered 0.5-3mA biphasic current stimulation. The SoC is powered wirelessly and bidirectional data telemetry is realized through the same pair of coils in 13.56MHz.
URI: http://hdl.handle.net/11536/147219
期刊: 2017 SYMPOSIUM ON VLSI CIRCUITS
顯示於類別:會議論文