<link rel="stylesheet" href="styles.cf390c6bfbe7555f.css"> Skip to main content

Mathematical Derivation, Circuits Design and Clinical Experiments of Measuring Blood Flow Volume (BFV) at Arteriovenous Fistula (AVF) of Hemodialysis (HD) Patients Using a Newly-Developed Photoplethysmography (PPG) Sensor

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

DOI

Abstract

Mathematical derivation of calculating blood flow volume (BFV) at arteriovenous fistula (AVF) using a newly-developed photoplethysmography (PPG) sensor is presented in this work. Also, the readout circuit of the PPG sensor intended to increase the signal-noise ratio (S/NR) is designed and presented in this work. The designed PPG sensor equipped with derived mathematical equations shows high correlation (R-2 = 0.7563) and low error (RMSE = 212 ml/min) as compared to the gold standard.