dc.description.abstractA practical approach to reduce the interferences of biochemicals and hematocrit ratio (Hct%) in the determination of whole blood glucose using multiple screen-printed carbon electrode (SPCE) test strips is described. SPCE test strips with and without glucose oxidase [i.e., GOD(+)-SPCEs and GOD(-)-SPCEs] were used and the chronoamperometric currents of test glucose solutions with various spiked uric acid concentrations and Hct% were measured. By establishing the interference relationships between glucose concentrations and uric acid concentrations as well as Hct% values and with appropriate corrections, the whole blood glucose determinations could be made to be more accurate and comparable to those determined by the reference YSI method. Specifically, the use of the Delta I value, i.e., the current difference between GOD(+)-SPCE and GOD(-)-SPCE measurements, would reduce most of the uric acid/biochemical interferences. An interpolation method was also established to correct for the glucose determinations with Hct% interferences. The Hct% corrections using the interpolation method are especially important and necessary for those blood samples with glucose concentrations higher than 110 mg dL(-1) and Hct% values lower than 35%. This approach should also be applicable to other biochemical determinations using similar electrochemical techniques.en_US
dc.language.isoen_USen_US
dc.subjectscreen-printed carbon electrodesen_US
dc.subjectelectrochemical biosensoren_US
dc.subjectglucoseen_US
dc.subjectinterferenceen_US
dc.subjecthematocrit ratioen_US
dc.subjecturic aciden_US
dc.titleReduction of the interferences of biochemicals and hematocrit ratio on the determination of whole blood glucose using multiple screen-printed carbon electrode test stripsen_US
dc.typeArticleen_US
dc.identifier.doi10.1007/s00216-007-1521-7en_US
dc.identifier.journalANALYTICAL AND BIOANALYTICAL CHEMISTRYen_US
dc.citation.volume389en_US
dc.citation.issue5en_US
dc.citation.spage1623en_US
dc.citation.epage1631en_US
dc.contributor.department生物科技學系zh_TW
dc.contributor.department電控工程研究所zh_TW
dc.contributor.departmentDepartment of Biological Science and Technologyen_US
dc.contributor.departmentInstitute of Electrical and Control Engineeringen_US
dc.identifier.wosnumberWOS:000250537100033-
dc.citation.woscount3-
顯示於類別:期刊論文


文件中的檔案:

  1. 000250537100033.pdf

若為 zip 檔案,請下載檔案解壓縮後,用瀏覽器開啟資料夾中的 index.html 瀏覽全文。