dc.description.abstractBackground: Home glucose monitoring system is increasingly recognized as an important tool for glycemic control. We evaluated the clinical performance of the eBsensor glucose monitoring system. Methods: Fingertip capillary blood glucose concentrations from 282 subjects were measured using eBsensor glucose monitoring system and compared against predicate devices and the Yellow Springs Instruments (YSI) 2300 blood glucose analyzer. Accuracy and precision of the eBsensor glucose monitoring system were assessed using several methods. The comparative study between the eBsensor and 2 currently marketed monitoring systems was performed. Results: The 282 eBsensor readings covered a wide range from 2.6 to 24.4 mmol/l. Deming regression and Pearson correlation analyses showed a linear relationship between the eBsensor readings and the YSI reference method (eBsensor=0.9496 YSI+0.4127 mmol/l; r=0.98). Error Grid analysis demonstrated that 100% of the eBsensor readings in clinically acceptable zones A and B. The CVs for the 6 lots of strips were within the satisfactory interval (< 6%). The comparative study showed that the eBsensor readings correlated well with the OneTouch Ultra values (r=0.97) and the Glucocard II values (r=0.97). Conclusions: eBsensor is a reliable glucose monitoring system which provides high accurate and precise glucose readings over a wide range of glucose concentrations. (c) 2006 Elsevier B.V. All rights reserved.en_US
dc.language.isoen_USen_US
dc.subjectdiabetesen_US
dc.subjectglucose monitoring systemen_US
dc.subjectClarke error grid analysisen_US
dc.subjectglucose meteren_US
dc.subjectstripen_US
dc.subjectself-monitoringen_US
dc.subjectaccuracyen_US
dc.subjectprecisionen_US
dc.subjectelectrochemicalen_US
dc.titleClinical evaluation of the eBsensor hand-held blood glucose monitoring systemen_US
dc.typeArticleen_US
dc.identifier.doi10.1016/j.cca.2006.09.023en_US
dc.identifier.journalCLINICA CHIMICA ACTAen_US
dc.citation.volume377en_US
dc.citation.issue1-2en_US
dc.citation.spage170en_US
dc.citation.epage173en_US
dc.contributor.department生物科技學系zh_TW
dc.contributor.departmentDepartment of Biological Science and Technologyen_US
dc.identifier.wosnumberWOS:000243675000023-
dc.citation.woscount3-
顯示於類別:期刊論文


文件中的檔案:

  1. 000243675000023.pdf

若為 zip 檔案,請下載檔案解壓縮後,用瀏覽器開啟資料夾中的 index.html 瀏覽全文。