標題: | Computations of the Signal-to-Noise Ratio for Medical X-Ray Imagers |
作者: | Lo, Yu-Chen Huang, Kuan-Chieh Yang, Kei-Hsiung 光電系統研究所 影像與生醫光電研究所 Institute of Photonic System Institute of Imaging and Biomedical Photonics |
公開日期: | 1-八月-2014 |
摘要: | If the signal in a medical X-ray imaging detector is formed by integrating the scintillation (or electrical) pulses, rather than by counting them, the signal-to-noise ratio is reduced by a factor which depends on the pulse height distribution. Formulas are expanded from Swank\'s publication to derive the signal-to-noise ratio for any given optical or electrical pulse-size distribution. A new quantity called noise-equivalent absorption is defined which bears a simple relationship to the signal-to-noise ratio of integrated X-ray imaging as the quantum absorption to the signal-to-noise ratio of photon-counting X-ray imaging. |
URI: | http://dx.doi.org/10.6122/CJP.52.1277 http://hdl.handle.net/11536/123992 |
ISSN: | 0577-9073 |
DOI: | 10.6122/CJP.52.1277 |
期刊: | CHINESE JOURNAL OF PHYSICS |
Volume: | 52 |
起始頁: | 1277 |
結束頁: | 1287 |
顯示於類別: | 期刊論文 |