標題: | A compact 3D-printed interface for coupling open digital microchips with Venturi easy ambient sonic-spray ionization mass spectrometry |
作者: | Hu, Jie-Bi Chen, Ting-Ru Chang, Chia-Hsien Cheng, Ji-Yen Chen, Yu-Chie Urban, Pawel L. 應用化學系 應用化學系分子科學碩博班 Department of Applied Chemistry Institute of Molecular science |
公開日期: | 1-Jan-2015 |
摘要: | Digital microfluidics (DMF) based on the electrowetting-on-dielectric phenomenon is a convenient way of handling microlitre-volume aliquots of solutions prior to analysis. Although it was shown to be compatible with on-line mass spectrometric detection, due to numerous technical obstacles, the implementation of DMF in conjunction with MS is still beyond the reach of many analytical laboratories. Here we present a facile method for coupling open DMF microchips to mass spectrometers using Venturi easy ambient sonic-spray ionization operated at atmospheric pressure. The proposed interface comprises a 3D-printed body that can easily be "clipped" at the inlet of a standard mass spectrometer. The accessory features all the necessary connections for an open-architecture DMF microchip with T-shaped electrode arrangement, thermostatting of the microchip, purification of air (to prevent accidental contamination of the microchip), a Venturi pump, and two microfluidic pumps to facilitate transfer of samples and reagents onto the microchip. The system also incorporates a touch-screen panel and remote control for user-friendly operation. It is based on the use of popular open-source electronic modules, and can readily be assembled at low expense. |
URI: | http://dx.doi.org/10.1039/c4an02220c http://hdl.handle.net/11536/124386 |
ISSN: | 0003-2654 |
DOI: | 10.1039/c4an02220c |
期刊: | ANALYST |
Volume: | 140 |
起始頁: | 1495 |
結束頁: | 1501 |
Appears in Collections: | Articles |