標題: | A versatile QCM matrix system for online and high-throughput bio-sensing |
作者: | Huang, GS Wang, MT Hong, MY 材料科學與工程學系奈米科技碩博班 Graduate Program of Nanotechnology , Department of Materials Science and Engineering |
公開日期: | 2006 |
摘要: | A 3 x 3 quartz crystal microbalance (QCM) sensor matrix, fabricated on an A-T cut quartz crystal, has the ability to detect online a variety of labeled DNA samples in a parallel and comparative fashion. The QCM matrix was equipped with a single oscillator circuit, which activated only one QCM at any given time, and was controlled by programmable time-shared electronic relays. The gold electrode had a diameter of 0.8 mm and operated at a fundamental resonating frequency of 40MHz; the dimensions of the matrix were 1.2 cm x 1.2 cm. The sensitivity of an individual QCM was in the pictogram regime. Selected QCMs were coated with either streptavidin or the anti-DIG antibody; the specificity of their detections was monitored using various concentrations of samples of biotin- and DIG-labeled DNA. The basic design of the QCM matrix is readily expandable, without any conceivable difficulties, in both geometry and circuitry. |
URI: | http://hdl.handle.net/11536/12871 http://dx.doi.org/10.1039/b515722f |
ISSN: | 0003-2654 |
DOI: | 10.1039/b515722f |
期刊: | ANALYST |
Volume: | 131 |
Issue: | 3 |
起始頁: | 382 |
結束頁: | 387 |
顯示於類別: | 期刊論文 |