標題: | A cost-effective fabrication of iridium oxide films as biocompatible electrostimulation electrodes for neural interface applications |
作者: | Chen, Yong-Min Chung, Tsai-Wei Wu, Pu-Wei Chen, Po-Chun 材料科學與工程學系 Department of Materials Science and Engineering |
關鍵字: | Neurostimulation;Bio-interface;Biocompatible electrostimulation electrode;Iridium oxide;Chemical bath deposition |
公開日期: | 25-一月-2017 |
摘要: | Electrostimulation medical devices for neural diseases require electroactive and biocompatible interface materials to transmit signals from electrodes to targeting tissues. Iridium oxide is an attractive ceramic material for neurostimulation electrodes due to its desirable stability and biocompatibility. In this study, we developed a cost effective and thickness controllable process to fabricate iridium oxide film by chemical bath deposition (CIROF). Surface morphology, crystallinity, roughness, hydrophilicity, and charge storage capacity as well as biocompatibility of the CIROFs with different thicknesses were analyzed. Accordingly, the CIROFs present rough and hydrophilic surfaces for cell attachment and good biocompatibility for cell viability. In addition, the charge storage capacity (CSC) of the as-deposited and the annealed CIROFs were 78.29 mC/cm(2) and 71.76 mC/cm(2), respectively. Consequently, these results show advantages that make the CIROF a promising neurostimulation electrode material for implantable medical devices in neural systems. (C) 2016 Elsevier B.V. All rights reserved. |
URI: | http://dx.doi.org/10.1016/j.jallcom.2016.09.064 http://hdl.handle.net/11536/132593 |
ISSN: | 0925-8388 |
DOI: | 10.1016/j.jallcom.2016.09.064 |
期刊: | JOURNAL OF ALLOYS AND COMPOUNDS |
Volume: | 692 |
起始頁: | 339 |
結束頁: | 345 |
顯示於類別: | 期刊論文 |