SU-8 resist was used as a core/cladding waveguide material to fabricate a Mach-Zehnder interferometer (MZI) for biochemical sensing. The refractive index of the SU-8 resist was fine-tuned with a Delta n of 0.004 for single-mode transmission. The UV lithography processes of the SU-8 resist were also optimized to pattern the high resolution (1 mu m) and high aspect ratio (A(R) = 6) Y-branch structure of the optical interferometer. Optical measurements reveal that the SU-8 MZI chip can efficiently transmit the NIR laser (lambda = 13 10 mu) with a total loss less than 6 dB. When one branch of the MZI is in contact with the analyte, the interfered intensity stabilizes after the soaking time exceeds 5 min. NaCl solution with a concentration of 10(-9) g/l can be detected using the SU-8 MZI chip. In the future, the polymer MZI chip can be mass-produced by molding process (or the LIGA process). The low-cost, label-free, real-time and high-sensitivity MZI chip will benefit many applications related to biological, environmental and industrial detection. (c) 2005 Elsevier B.V. All rights reserved.