标题: 使用多感测器系统实作手指颤抖分析并应用于神经退化疾病
Using Multi-Sensor System for Measuring Finger Trembling, and Applied to Neurodegenerative Disease
作者: 陈咨频
荆宇泰
Chen, Tzu-Pin
Ching, Yu-Tai
生医工程研究所
关键字: Tremogram;Leap Motion;加速度计;陀螺仪;手指颤抖分析;Tremogram;Leap Motion;Accelerometer;Gyroscope;Finger Trembling Analysis
公开日期: 2017
摘要: 手指颤抖对于神经退化疾病来说是一个很重要的特征,透过手指颤抖的情形可以帮助医师判断疾病。Tremogram 是目前医院所使用的侦测仪器,其原理是使用加速规来做侦测,并且将获得的数据经由快速傅立叶转换来得知手指颤抖的频率。但是 Tremogram 目前用的是单一且单轴的加速规,对于病患来说,无法一次就做全面性的手指颤抖检查。

之前有论文实作 Leap Motion 与 Tremogram 的比较。但是经过 Leap Motion 版本的更新后,敏感度较先前的版本低。因此本论文主要使用多感测器系统的G-sensor来做手指颤抖的资料搜集,并且设计演算法来衍伸出更多的资料分析。

论文中分析了三位介于 24 ~ 28 岁的正常人及两位有手指颤抖病征的病人。由实验结果发现,使用多感测器系统得到的分析结果很接近目前医院使用的 Tremogram,而且还可以一次得到五支手指头在空间中三维的资料,也藉着不同的演算法将颤抖程度量化。本论文成功的利用较低成本的机器实作出比 Tremogram 更多元、并且比 Leap Motion 敏感度更高的分析结果。期待本论文所实作出来的系统可以进一步应用在临床医学上。
Finger trembling is an important feature for neurodegenerative disease, through the trembling of the fingers can help the doctor to determine the disease. Tremogram is currently used by the hospital detection equipment. Then transfer the data to frequency domain by using fast Fourier transform to get the frequency of finger trembling. However, Tremogram is currently using a single and uniaxial acceleration gauge, which make patients, can not do a comprehensive finger trembling check.

There is a research comparison of Leap Motion and Tremogram. But after update the new version of Leap Motion.The sensitivity is lower then the previous version. Therefore, the purpose of my research is to explores the other tools to do finger trembling data collection, and design algorithms to analysis the data.

There are three normal people and two patients with finger tremors as experimental subjects in my research. From the experiment results, it is found that not only the analysis result obtained by using the multi-sensor system is very close to Tremogram, but also my system can quantize the amplitude of all fingers trembling on 3-axis at once. My research succeeded in using lower-cost machines to produce more diverse information than Tremogram and more sensitive than Leap Motion. It is expected that the system can applied in the clinical in the future.