dc.description.abstractThe 6-minute walking test (6MWT) is the test most commonly used to evaluate cardiopulmonary function in patients with respiratory or heart disease. However, there was previously no integrated monitoring system available to simultaneously record both the real-time cardiopulmonary physiological parameters and the walking information (i.e., walking distance, speed, and acceleration) during the 6MWT. In this study, then, a wearable and wireless multi-parameter monitoring system was proposed to simultaneously monitor oxygen saturation (SpO(2)), heart rhythm, and the walking information during the 6MWT. A multi parameter detection algorithm was also designed to estimate the heart rate effectively. The results of the study indicate that this system was able to reveal the dynamic changes and differences in walking speed and acceleration during the 6MWT. As such, the system has the potential to provide a more integrated approach to monitoring cardiopulmonary parameters and walking information simultaneously during the 6MWT. The proposed system warrants further investigation as an assistive assessment tool in evaluating cardiopulmonary function and may be widely applied in cardiopulmonary-related and sports medicine applications in the future. (C) 2017 IPEM. Published by Elsevier Ltd. All rights reserved.en_US
dc.language.isoen_USen_US
dc.subjectSix-minute walking testen_US
dc.subjectCardiopulmonary functionen_US
dc.subjectWalking informationen_US
dc.subjectHeart rateen_US
dc.titleDesign of wearable and wireless multi-parameter monitoring system for evaluating cardiopulmonary functionen_US
dc.typeArticleen_US
dc.identifier.doi10.1016/j.medengphy.2017.06.009en_US
dc.identifier.journalMEDICAL ENGINEERING & PHYSICSen_US
dc.citation.volume47en_US
dc.citation.spage144en_US
dc.citation.epage150en_US
dc.contributor.department影像與生醫光電研究所zh_TW
dc.contributor.departmentInstitute of Imaging and Biomedical Photonicsen_US
dc.identifier.wosnumberWOS:000409294500016en_US