標題: Kinetic study of continuous liquid-liquid extraction of wine with real-time detection
作者: Yang, Hui-Hsien
Dutkiewicz, Ewelina P.
Urban, Pawel L.
應用化學系
Department of Applied Chemistry
關鍵字: Automation;Continuous liquid/liquid extraction;Hyphenated techniques;Mass spectrometry
公開日期: 30-十一月-2018
摘要: Kinetic optimization of continuous liquid-liquid extraction (CLLE) can shorten sample preparation times and reduce losses of labile or volatile analytes. Here, we coupled a downscaled CLLE apparatus with atmospheric pressure chemical ionization interface of triple quadrupole mass spectrometer. Real-time sampling was guided by an Arduino-based programmable logic controller. The recorded datasets were processed to compute the extraction rate constants for the target analytes. The extraction time in subsequent on-line experiments was set to 180 min as a compromise between the reduction of the analysis time and maximizing its yield. Interestingly, off-line analysis of the extract produced different results than on-line analysis pointing to the immanent degradation of the collected extract aliquots. Next, we implemented this hyphenated system in the analysis of red wine samples, which were stored during different periods of time after opening the bottle. The results reveal differences in the depletion of the volatile wine components during storage. (C) 2018 Elsevier B.V. All rights reserved.
URI: http://dx.doi.org/10.1016/j.aca.2018.06.072
http://hdl.handle.net/11536/148082
ISSN: 0003-2670
DOI: 10.1016/j.aca.2018.06.072
期刊: ANALYTICA CHIMICA ACTA
Volume: 1034
起始頁: 85
結束頁: 91
顯示於類別:期刊論文