標題: | Thermopneumatic suction integrated microfluidic blood analysis system |
作者: | Yang, Chiao-Hsun Hsieh, Yu-Ling Tsou, Ping-Hsien Li, Bor-Ran 交大名義發表 生物科技學系 生醫工程研究所 National Chiao Tung University Department of Biological Science and Technology Institute of Biomedical Engineering |
公開日期: | 7-三月-2019 |
摘要: | Blood tests provide crucial diagnostic information regarding several diseases. A key factor that affects the precision and accuracy of blood tests is the interference of red blood cells; however, the conventional methods of blood separation are often complicated and time consuming. In this study, we devised a simple but high-efficiency blood separation system on a self-strained microfluidic device that separates 99.7 +/- 0.3% of the plasma in only 6 min. Parameters, such as flow rate, design of the filter trench, and the relative positions of the filter trench and channel, were optimized through microscopic monitoring. Moreover, this air difference-driven device uses a cost-effective and easy-to-use heater device that creates a low-pressure environment in the microchannel within minutes. With the aforementioned advantages, this blood separation device could be another platform choice for point-of-care testing. |
URI: | http://dx.doi.org/10.1371/journal.pone.0208676 http://hdl.handle.net/11536/148971 |
ISSN: | 1932-6203 |
DOI: | 10.1371/journal.pone.0208676 |
期刊: | PLOS ONE |
Volume: | 14 |
顯示於類別: | 期刊論文 |