標題: Thermopneumatic suction integrated microfluidic blood analysis system
作者: Yang, Chiao-Hsun
Hsieh, Yu-Ling
Tsou, Ping-Hsien
Li, Bor-Ran
交大名義發表
生物科技學系
生醫工程研究所
National Chiao Tung University
Department of Biological Science and Technology
Institute of Biomedical Engineering
公開日期: 7-三月-2019
摘要: Blood tests provide crucial diagnostic information regarding several diseases. A key factor that affects the precision and accuracy of blood tests is the interference of red blood cells; however, the conventional methods of blood separation are often complicated and time consuming. In this study, we devised a simple but high-efficiency blood separation system on a self-strained microfluidic device that separates 99.7 +/- 0.3% of the plasma in only 6 min. Parameters, such as flow rate, design of the filter trench, and the relative positions of the filter trench and channel, were optimized through microscopic monitoring. Moreover, this air difference-driven device uses a cost-effective and easy-to-use heater device that creates a low-pressure environment in the microchannel within minutes. With the aforementioned advantages, this blood separation device could be another platform choice for point-of-care testing.
URI: http://dx.doi.org/10.1371/journal.pone.0208676
http://hdl.handle.net/11536/148971
ISSN: 1932-6203
DOI: 10.1371/journal.pone.0208676
期刊: PLOS ONE
Volume: 14
顯示於類別:期刊論文