Title: | Biomechanical investigation of dynamic hip screw and wire fixation on an unstable intertrochanteric fracture |
Authors: | Wu, Hsu-Fu Chang, Chih-Han Wang, Gwo-Jaw Lai, Kuo-An Chen, Chung-Hwan 交大名義發表 National Chiao Tung University |
Keywords: | Dynamic hip screw;Intertrochanteric hip fracture;Lesser trochanteric fragment;Wire;Unstable type |
Issue Date: | 24-Apr-2019 |
Abstract: | BackgroundAlthough use of a dynamic hip screw (DHS) for stable intertrochanteric hip fracture fixation has been successfully applied in fracture healing for more than 20years, DHS fixation on unstable intertrochanteric fractures still has a high failure rate, especially in patients with osteoporosis. Although the wire fixation is usually incorporated with orthopedic device to treat fracture, the wiring techniques are developed through experiences. Thus, this study is objective to investigate the biomechanical property of different wire fixation methods incorporated with DHS system to provide the lesser trochanter fragment stable fixation on osteoporotic TypeA2.1 fracture for enhancing stability after bone reduction.ResultsSawbone testing results demonstrated higher maximum load, stiffness, and energy in a DHS with wire fixation compared with DHS fixation only. In static biomechanical testing of a cadaver femur, we compared the stiffness of five fixation models and then tested a fatigue failure model in cycle loading with DHS fixation only. Wiring fixation can enhance stability and the cut-out failure model in the fatigue test was identical to the clinical failure model.ConclusionsLesser trochanteric fragment fixation is a crucial concern in the stability of an A2.1 unstable fracture, and the combination of a wiring technique with a DHS seems beneficial for achieving better stability. The addition of an antirotational greater trochanter is likely to enhance stability through wiring of the greater trochanter. |
URI: | http://dx.doi.org/10.1186/s12938-019-0663-0 http://hdl.handle.net/11536/151918 |
ISSN: | 1475-925X |
DOI: | 10.1186/s12938-019-0663-0 |
Journal: | BIOMEDICAL ENGINEERING ONLINE |
Volume: | 18 |
Begin Page: | 0 |
End Page: | 0 |
Appears in Collections: | Articles |