標題: A mass correction method for the aerosol particle mass analyzer to measure the particle mass of sub-50 nm nanoparticles
作者: Liao, Bo-Xi
Gong, Wen-Cheng
Li, Ziyi
Tsai, Chuen-Jinn
環境工程研究所
Institute of Environmental Engineering
關鍵字: Kihong Park
公開日期: 1-一月-1970
摘要: The particle mass measurement instrument of the aerosol particle mass analyzer (APM) tends to underestimate the particle mass of sub-50 nm nanoparticles, limiting the APM mass measurement to larger particles. In the present work, a mass correction method was developed based on a calibration reference to correct the APM mass underestimation according to a measured mass-based dimensionless parameter The correction was tested using laboratory generated silver nanoparticles and ambient nanoparticles. It was found that the measured effective density variations of silver nanoparticles due to different extent of APM mass underestimation at different operating conditions were reduced by the correction method with a significant improvement of R-squared from 0.40 to 0.87. In addition, about 20% increase in the effective density from 0.91 to 1.09 g/cm(3) was observed for 30 nm ambient wet nanoparticles using the correction method. The difference between the raw and corrected effective densities is due to the mass underestimation by the APM itself, which should be corrected in the data analysis program of the APM. Copyright (c) 2019 American Association for Aerosol Research
URI: http://dx.doi.org/10.1080/02786826.2019.1626345
http://hdl.handle.net/11536/152154
ISSN: 0278-6826
DOI: 10.1080/02786826.2019.1626345
期刊: AEROSOL SCIENCE AND TECHNOLOGY
起始頁: 0
結束頁: 0
顯示於類別:期刊論文