標題: Design of a Wearable 12-Lead Noncontact Electrocardiogram Monitoring System
作者: Hsu, Chien-Chin
Lin, Bor-Shing
He, Ke-Yi
Lin, Bor-Shyh
影像與生醫光電研究所
Institute of Imaging and Biomedical Photonics
關鍵字: electrocardiogram;conductive gels;noncontact electrode;myocardial ischemia;pacemaker;ventricular premature contraction
公開日期: 1-四月-2019
摘要: A standard 12-lead electrocardiogram (ECG) is an important tool in the diagnosis of heart diseases. Here, Ag/AgCl electrodes with conductive gels are usually used in a 12-lead ECG system to access biopotentials. However, using Ag/AgCl electrodes with conductive gels might be inconvenient in a prehospital setting. In previous studies, several dry electrodes have been developed to improve this issue. However, these dry electrodes have contact with the skin directly, and they might be still unsuitable for patients with wounds. In this study, a wearable 12-lead electrocardiogram monitoring system was proposed to improve the above issue. Here, novel noncontact electrodes were also designed to access biopotentials without contact with the skin directly. Moreover, by using the mechanical design, this system allows the user to easily wear and take off the device and to adjust the locations of the noncontact electrodes. The experimental results showed that the proposed system could exactly provide a good ECG signal quality even while walking and could detect the ECG features of the patients with myocardial ischemia, installation pacemaker, and ventricular premature contraction.
URI: http://dx.doi.org/10.3390/s19071509
http://hdl.handle.net/11536/152359
ISSN: 1424-8220
DOI: 10.3390/s19071509
期刊: SENSORS
Volume: 19
Issue: 7
起始頁: 0
結束頁: 0
顯示於類別:期刊論文