<link rel="stylesheet" href="styles.cf390c6bfbe7555f.css"> Skip to main content

An antifouling peptide-based biosensor for determination of Streptococcus pneumonia markers in human serum

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

DOI

10.1016/j.bios.2019.111969

Abstract

We report a peptide-based sensor that involves a multivalent interaction with L-ascorbate 6-phosphate lactonase (UlaG), a protein marker of Streptococcus pneumonia. By integrating the antifouling feature of the sensor, we significantly improved the signal-to-noise ratio of UlaG detection. The antifouling surface was fabricated via electrodeposition using an equivalent mixture of 4-amino-N,N,N-trimethylanilinium and 4-aminobenzenesulfonate. This antifouling layer not only effectively reduces the non-specific adsorption on the biosensor but also decreases the charge transfer resistance (R-et) of the screen-printed carbon electrode. The aniline-modified 57 peptide, an UlaG-binding peptide, was pre-synthesized and further electrochemically modified to bind onto the antifouling layer. Bio-electrochemical analysis confirms that the antifouling 57-peptide sensor binds strongly to the UlaG with a dissociation constant (K-d) = 0.5 nM. This strong interaction can be attributed to a multivalent interaction between the biosensor and the heximeric form of UlaG. To demonstrate the potential for clinical application, further detection of Streptococcus pneumonia from 50 to 5 x 10(4) CFU/mL were successfully performed in 25% human serum.

Description