dc.description.abstractA novel multifilter PM(10)-PM(2.5) sampler (MFPPS) that enables the collection of four PM(10) and four PM(2.5) samples simultaneously has been developed and tested. The MFPPS uses a PM(10) impactor as the inlet and operates at 33.4 L/min. After the inlet, the aerosol flow is divided half by a Y-type fitting. Half of the flow is directed into four PM(10) filter cassettes, while the other half is directed into four PM(2.5) filter cassettes after the aerosols are further classified by a PM(2.5) impactor. An active flow control system consisting of two mass flow controllers (MFCs), one for PM(10) and the other for PM(2.5), is used to fix the total flow rate of 16.7 L/min for four PM(10) or four PM(2.5) channels based on the ambient pressure and temperature. To ensure flow rate uniformity through each of the four PM(10) or four PM(2.5) filter cassettes, an orifice is assembled behind each of the filter cassettes to increase the pressure drop, such that the flow rates of eight sampling lines are nearly equal using just two MFCs. The MFPPS was calibrated in the laboratory for particle collection efficiency curves first. Then, the ambient PM concentrations were compared with those of other two collocated FRM samplers, the dichotomous PM(10) and the EPA WINS PM(2.5) sampler in the field study. Calibration results showed the cutoff aerodynamic diameters of the PM(10) and PM(2.5) impactors were 9.8 +/- 0.1 and 2.5 +/- 0.05 mu m, respectively. Field comparison results indicated PM(10) and PM(2.5) concentrations agreed well with the other two PM samplers.en_US
dc.language.isoen_USen_US
dc.titleA Novel Multifilter PM(10)-PM(2.5) Sampler (MFPPS)en_US
dc.typeArticleen_US
dc.identifier.doi10.1080/02786826.2011.602135en_US
dc.identifier.journalAEROSOL SCIENCE AND TECHNOLOGYen_US
dc.citation.volume45en_US
dc.citation.issue12en_US
dc.citation.spage1480en_US
dc.citation.epage1487en_US
dc.contributor.department環境工程研究所zh_TW
dc.contributor.departmentInstitute of Environmental Engineeringen_US
顯示於類別:期刊論文