標題: A calcium vapor cell for atomic spectroscopy
作者: Huang, MS
Lu, MH
Shy, JT
物理研究所
光電工程學系
Institute of Physics
Department of Photonics
公開日期: 1-十一月-2002
摘要: A calcium vapor cell used in laser spectroscopy usually suffers from calcium coating on the cell windows. To solve the problem, we add two mirrors in the cell to reflect the laser beam, then the calcium will coat on these mirrors instead of on the windows. Calcium coating degrades the reflectivity of the mirrors at the beginning, but the mirrors become calcium mirrors after a certain period of coating and their reflectivity will not change by further calcium coating. Using this cell in a conventional saturated-absorption spectroscopy, the observed saturated-absorption peak at 657 nm is about 2.4% of the probe beam power and its linewidth is below 300 kHz. Therefore, a portable secondary calcium-wavelength standard at 657 nm having an accuracy better than 1x10(-11) is possible. In addition, our cell has a continuous working time longer than 3 d and it can be easily extended to 10 d. (C) 2002 American Institute of Physics.
URI: http://dx.doi.org/10.1063/1.1511800
http://hdl.handle.net/11536/28426
ISSN: 0034-6748
DOI: 10.1063/1.1511800
期刊: REVIEW OF SCIENTIFIC INSTRUMENTS
Volume: 73
Issue: 11
起始頁: 3747
結束頁: 3749
顯示於類別:期刊論文


文件中的檔案:

  1. 000178818200004.pdf

若為 zip 檔案,請下載檔案解壓縮後,用瀏覽器開啟資料夾中的 index.html 瀏覽全文。