標題: Field study of the accuracy of two respirable sampling cyclones
作者: Tsai, CJ
Shiau, HG
Shih, TS
環境工程研究所
Institute of Environmental Engineering
公開日期: 1-十二月-1999
摘要: This study is to evaluate the sampling bias of a new cyclone (Tsai et al. 1999) and a 10 mm nylon cyclone under field sampling conditions. Both cyclones were operated at the optimum flow rate of 1.7 L/min. The Marple personal cascade impactor was used to measure the particle size distributions, which were used to calculate the reference RPM (Respirable Particulate Matter) concentrations according to the new ACGIH criteria. A wind tunnel was used in the field to ensure uniform particle concentration and wind speed across the tested cyclones and impactors in the working section. The wind speed, which was varied from 0.8 to 2.3 m/s, was kept constant in the wind tunnel. Near the packaging area of a glazing-making factory (mass median aerodynamic diameter, or MMAD: 11.1 mu m), the average ratio of the RPM concentrations measured by the cyclones to the standard RPM concentrations was found to be 1.20 and 1.10 for the 10 mm nylon and the new cyclone, respectively. Near the packaging area of a lead powder factory (MMAD: 4.82 mu m), the average ratio of the RPM measured by the cyclones to the reference RPM was found to be 1.13 and 1.02 for the 10 mm nylon and the new cyclone, respectively. Wind speed and total airborne particle concentration were found to have little effect on the accuracy of the two respirable cyclones. This field study showed that the new cyclone is more accurate than the 10 mm nylon cyclone based on the reference RPM.
URI: http://dx.doi.org/10.1080/027868299304011
http://hdl.handle.net/11536/30917
ISSN: 0278-6826
DOI: 10.1080/027868299304011
期刊: AEROSOL SCIENCE AND TECHNOLOGY
Volume: 31
Issue: 6
起始頁: 463
結束頁: 472
顯示於類別:期刊論文


文件中的檔案:

  1. 000083886600006.pdf

若為 zip 檔案,請下載檔案解壓縮後,用瀏覽器開啟資料夾中的 index.html 瀏覽全文。