標題: Novel Dry Polymer Foam Electrodes for Long-Term EEG Measurement
作者: Lin, Chin-Teng
Liao, Lun-De
Liu, Yu-Hang
Wang, I-Jan
Lin, Bor-Shyh
Chang, Jyh-Yeong
資訊工程學系
電信工程研究所
腦科學研究中心
Department of Computer Science
Institute of Communications Engineering
Brain Research Center
關鍵字: Biopotentials;conduction gel;dry electrode;electroencephalography;skin-electrode interfaces impedance
公開日期: 1-五月-2011
摘要: A novel dry foam-based electrode for long-term EEG measurement was proposed in this study. In general, the conventional wet electrodes are most frequently used for EEG measurement. However, they require skin preparation and conduction gels to reduce the skin-electrode contact impedance. The aforementioned procedures when wet electrodes were used usually make trouble to users easily. In order to overcome the aforesaid issues, a novel dry foam electrode, fabricated by electrically conductive polymer foam covered by a conductive fabric, was proposed. By using conductive fabric, which provides partly polarizable electric characteristic, our dry foam electrode exhibits both polarization and conductivity, and can be used to measure biopotentials without skin preparation and conduction gel. In addition, the foam substrate of our dry electrode allows a high geometric conformity between the electrode and irregular scalp surface to maintain low skin-electrode interface impedance, even under motion. The experimental results presented that the dry foam electrode performs better for long-term EEG measurement, and is practicable for daily life applications.
URI: http://dx.doi.org/10.1109/TBME.2010.2102353
http://hdl.handle.net/11536/8942
ISSN: 0018-9294
DOI: 10.1109/TBME.2010.2102353
期刊: IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
Volume: 58
Issue: 5
起始頁: 1200
結束頁: 1207
顯示於類別:期刊論文


文件中的檔案:

  1. 000289807300008.pdf

若為 zip 檔案,請下載檔案解壓縮後,用瀏覽器開啟資料夾中的 index.html 瀏覽全文。