<link rel="stylesheet" href="styles.cf390c6bfbe7555f.css"> Skip to main content

Direct field observation of the relative humidity effect on the beta-gauge readings

dc.citation.epage840en_US
dc.citation.issue6en_US
dc.citation.spage834en_US
dc.citation.volume56en_US
dc.citation.woscount3
dc.contributor.authorTsai, CJen_US
dc.contributor.authorChang, CTen_US
dc.contributor.authorHuang, CHen_US
dc.contributor.department環境工程研究所zh_TW
dc.contributor.departmentInstitute of Environmental Engineeringen_US
dc.date.accessioned2014-12-08T15:16:27Z
dc.date.available2014-12-08T15:16:27Z
dc.date.issued2006-06-01en_US
dc.description.abstractThe effect of ambient relative humidity (RH) on hourly particulate matter (PM10) readings of P-gauge monitors has been studied using two collocated monitors in the field. The inlet air of monitor 1 was conditioned with water vapor to increase its RH, whereas monitor 2 operated normally in ambient conditions. Experimental data showed that PM10 readings of monitor 1 were nearly the same as monitor 2, as long as the RH of its conditioned incoming air did not exceed similar to 80-85%. However, when the RH exceeded similar to 80-85%, PM10 readings of monitor 1 became higher than monitor 2, and the difference increased with increasing RH. The measurement of pressure drop across the filter was also conducted, and the data revealed that the increase of pressure drop per unit of PM,, concentration decreased when RH was higher than similar to 80-85%, as compared with the case when RH was lower than 80-85%. This is perhaps because of more porous structure of deposited particles in the beta-gauge monitor when RH is greater than similar to 80-85%. The theoretical calculation using an evaporation model and a thermodynamic model has been conducted to simulate the beta gauge readings. The results show that the theoretical PM10 concentrations using the evaporation model are in better agreement with the actual beta-gauge readings than those using the thermodynamic equilibrium model.en_US
dc.identifier.issn1047-3289en_US
dc.identifier.journalJOURNAL OF THE AIR & WASTE MANAGEMENT ASSOCIATIONen_US
dc.identifier.urihttps://ir.lib.nycu.edu.tw/handle/11536/12176
dc.identifier.wosnumberWOS:000238176000013
dc.language.isoen_USen_US
dc.titleDirect field observation of the relative humidity effect on the beta-gauge readingsen_US
dc.typeArticleen_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
000238176000013.pdf
Size:
1.15 MB
Format:
Adobe Portable Document Format

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed to upon submission
Description: