A high precision calibration of the nonlinear energy response at Daya Bay

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

DOI

10.1016/j.nima.2019.06.031

Abstract

A high precision calibration of the nonlinearity in the energy response of the Daya Bay Reactor Neutrino Experiment's antineutrino detectors is presented in detail. The energy nonlinearity originates from the particle-dependent light yield of the scintillator and charge-dependent electronics response. The nonlinearity model is constrained by.. calibration points from deployed and naturally occurring radioactive sources, the.. spectrum from B-12 decays, and a direct measurement of the electronics nonlinearity with a new flash analog-to-digital converter readout system. Less than 0.5% uncertainty in the energy nonlinearity calibration is achieved for positrons of kinetic energies greater than 1 MeV.