<link rel="stylesheet" href="styles.cf390c6bfbe7555f.css"> Skip to main content

Compact spectrometer system based on a gradient grating period guided-mode resonance filter

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

DOI

10.1364/OE.24.010972

Abstract

We demonstrate a compact spectrometer system by using a gradient grating period guided-mode resonance filter-mounted on a linear photodetector array-that exhibits spatially dependent resonance characteristics; a specific incident wavelength is reflected such that the underlying array pixels measure minimum intensity. A precalibrated transmission efficiency matrix is used to determine each pixel's transmission efficiency for specific wavelengths. Unknown spectral information can be calculated from the measured intensity. Grating periods of 250-388 nm in 2-nm increments are used in each 100-cycle period. Device length is 2.23 mm. Spectral range of 506-700 nm is measurable with 1-nm resolution. (C) 2016 Optical Society of America

Description

Keywords

Citation

Endorsement

Review

Supplemented By

Referenced By