<link rel="stylesheet" href="styles.cf390c6bfbe7555f.css"> Skip to main content

A Novel Wireless Photoplethysmography Blood-Flow Volume Sensor for Assessing Arteriovenous Fistula of Hemodialysis Patients

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

DOI

10.1109/TIE.2017.2701780

Abstract

A novel blood-flow volume (BFV) sensor is presented for assessing quality of arteriovenous fistula in hemodialysis patients via noninvasive reflectance-type photoplethysmography (PPG). BFV is nowadays in clinic practices evaluated using an ultrasound Doppler monitor, which is expensive, bulky, and can only be operated by well-trained medical personnels. This study is devoted to develop a low-cost, small-sized, portable, and easy-to-use PPG sensor that is capable of continuous measurement of BFV. New designs of front-end analog circuits, signal processing, and an intelligent neural network calibration method are employed to finally achieve high correlation (R-2 = 0.7176), as opposed to the ultrasound Doppler monitor, with the root-mean-squared errors successfully controlled under 289 ml/min.

Description

Citation

Endorsement

Review

Supplemented By

Referenced By