標題: A plug-and-measure diagnostic tool for flux, resolution and beam size of a soft X-ray beamline
作者: Wei, Ching-Hsuan
Lin, Hong-Cheu
Lin, Shang-Wei
Yuh, Jih-Young
材料科學與工程學系
Department of Materials Science and Engineering
關鍵字: windowless ion chamber;soft X-ray beamline;gas phase absorption;dynamic beam size monitor
公開日期: 1-Sep-2015
摘要: A compact multi-functional diagnostic tool has been installed for commissioning beamlines at the Taiwan Light Source. For a photon beam, the instrument can measure the photon flux, energy resolution and beam size, consecutively with a photodiode or gold mesh, a windowless gas cell and a movable fluorescence screen viewport. Two gratings with ruling densities of 350 and 1000 lines mm(-1) and dispersing photons of energies from 80 to 1200 eV were calibrated with a photon flux of 10(11) photon s(-1) at slit openings of 50 mu m x 50 mu m; a maximum resolving power of greater than 10000 was certified with an absorption spectra of gaseous samples. Pressure differences over four orders of magnitude were achieved between the ion chamber and the flux measurement chamber with a single capillary differential pumping stage. A sequence of beam profiles was measured by moving continuously in the vicinity of the nominal focal positions. This tool is useful in commissioning or trouble-shooting at most beamlines in a synchrotron facility.
URI: http://dx.doi.org/10.1107/S1600577515011923
http://hdl.handle.net/11536/128144
ISSN: 1600-5775
DOI: 10.1107/S1600577515011923
期刊: JOURNAL OF SYNCHROTRON RADIATION
Volume: 22
起始頁: 1289
結束頁: 1292
Appears in Collections:Articles


Files in This Item:

  1. ae2dc1d787ea6284d39ca48630c40e58.pdf

If it is a zip file, please download the file and unzip it, then open index.html in a browser to view the full text content.