標題: | Integration of a guided-mode resonance filter with microposts for in-cell protein detection |
作者: | Tu, Yi-Kai Tsai, Meng-Zhe Lee, I-Chin Hsu, Hsin-Yun Huang, Cheng-Sheng 機械工程學系 應用化學系 Department of Mechanical Engineering Department of Applied Chemistry |
公開日期: | 2016 |
摘要: | We present an integrated microfluidic system consisting of a label-free biosensor of a guided-mode resonance filter (GMRF) and a microfluidic channel with a micropost filter. The GMRF was fabricated through replica molding using an ultraviolet-curable polymer and a plastic substrate. An array of microposts (a diameter and height of 26.5 and 56 mu m, respectively, and a spacing between 7.5 and 9.5 mu m), fabricated on a silicon substrate through photolithography, was used as the filter. A double-sided tape was used to laminate the GMRF and a microfluidic chip such that the integrated device provides two functions: filtration of the cell debris and quantification of the in-cell protein concentration. By measuring the changes in the resonant wavelength from the GMRF, the detection of beta-actin in an unprocessed lysed cell sample was demonstrated; the cell debris was separated using the micropost filter to prevent false measurement. |
URI: | http://dx.doi.org/10.1039/c6an00023a http://hdl.handle.net/11536/133984 |
ISSN: | 0003-2654 |
DOI: | 10.1039/c6an00023a |
期刊: | ANALYST |
Volume: | 141 |
Issue: | 13 |
起始頁: | 4189 |
結束頁: | 4195 |
顯示於類別: | 期刊論文 |